Diaphragm-type vacuum pumps are often ideally suited for a
variety of medical devices. Unfortunately, flow pulsation can offer a challenge
to developers, specifically in flow measurement and control. This article will
focus on some simple techniques for reliable and repeatable measurement of flow
and pressure, particularly where the pumped media is air or an otherwise
compressible gas.
Diaphragm-type vacuum pumps and compressors are well suited
to a variety of medical device applications due to their ability to operate
efficiently over a broad range of flows and pressures while keeping the fluid
path completely sealed. A downside is flow pulsation—a limitation
characteristic of most reciprocating pumps—which can create challenges for
systems integrators in terms of flow measurement and control.
 |
| Figure 1: Pressure test setup |
A proper test setup for evaluation flow and pressure
performance for a pump working against a compression load is schematically
depicted in Figure 1. For flow measurement, a conventional rotameter is
depicted. Rotameters are simple, reliable and reasonably accurate (plus/minus a
few percent) devices that typically utilize a ball, or “float,” that rises in a
tapered tube as the flow increases. As reliable as rotameters can be however,
to produce accurate results, they require smooth flow. They are also typically
calibrated for operation at standard conditions. The test set-up depicted
achieves the former by including a rigid volume, or “tank,” of sufficient size
to damp out the inherent pulsations in flow produced by the compressor.
Note that the tank is positioned downstream of the loading
valve so that it is not subjected to the same pressure load as the compressor
under test. This ensures that the tank itself need not be rated to withstand a
high burst pressure in order to be safely used in most application. Sizing of
the tank is another consideration. Too much pulsation in the flow stream will
show up as apparent vibration, or bouncing, of the rotameter's float, and this
is an indication that the observed flow cannot be relied on as accurate.
As a rule of thumb, a tank volume of about one third the
maximum flow being measured—in terms of magnitude—should be sufficient. For
example, if the pump's free flow is about 30 LPM then a 10 Liter tank should
provide a good starting point. Another approach is to start with a relatively
small tank and increase its size until no change in the measured flow can be
observed. At the other end of the spectrum, a tank much larger than necessary
will certainly provide accurate results, but response time may become an issue.
Given that some small but finite amount of pressure will be required to raise
the float and move the gas through the connective tubing and fittings, there
will be some slight compression of the gas within the tank and even small
changes in tank pressure will take time to re-stabilize, particularly where the
measured flows are relatively small.
Once tank volume is determined, accuracy of the rotameter
readings can usually be relied upon over the full range of applied loads since
the added load—applied via a restrictive or throttling device (in this example,
a simple inline valve, or “Variable Restriction”)—will act to further damp out
flow pulsation.
Another important aspect of test set-up that can be seen in
the schematic is that the rotameter is the last element in the flow path. This
is necessary in order to keep one end of the device open to ambient, thereby ensuring
operation under essentially standard conditions.
 |
| Figure 2: Vacuum test setup |
A detail not shown in the schematic, but in some instances
helpful, is the introduction of a damping element between the pressure gauge
and the main flow path. In some cases, large pulsations in pump output can show
up as oscillations in the pressure gauge reading. Since the leg leading to the
gauge is essentially a dead end with relatively little internal volume, a
passive restrictive element can be introduced to stabilize gauge readout.
Figure 2 depicts the corresponding set-up for measuring pump
performance when subjected to loading on the inlet—vacuum load. Note that in
this case the rotameter's inlet is open to atmosphere. Again, this ensures that
the device will be operating at relatively standard conditions, and therefore
return the most accurate results.
A final consideration for test set-up is the selection and
sizing of connective tubing and fittings. These components should be selected so
as to present as little pressure drop in and of themselves as is practical. A good test is to check for minimal pressure
(or vacuum) reading at the gauge when the pump is running under no applied
load—where the variable restriction is completely open.
William Fleming is the president of Dynaflo, a supplier of smaller,
quieter, and more efficient vacuum pumps and compressors for the medical device
industry. He can be reached at 610-200-8017, x7011 or wtf@dynaflopumps.com.